SPECIFICATIONS
Physical
|
|
Component
|
Dimensions (W x D x H)
|
|
|
TIGRIS DTS Analyzer
|
69 x 36 x 72 in (175.3 x 91.4 x 183 cm)
|
|
|
Computer Workstation
|
33 x 36 in (84 x 91.4 cm)
|
|
|
UPS
|
7 x 8.8 x 17.1 in (18 x 22 x 44 cm)
|
|
|
Vacuum Pump
|
12 x 16.5 x 14 in (31 x 42 x 36 cm)
|
Weight
|
|
TIGRIS DTS Analyzer
|
1291 lbs (581 kg)
|
|
|
Computer Workstation
|
85 lbs (38 kg)
|
|
|
UPS
|
102 lbs (46 kg)
|
|
|
Vacuum Pump
|
35 lbs (16 kg)
|
Clearance Requirements
|
|
TIGRIS DTS Analyzer
|
96 in (244 cm) vertical clearance
|
|
|
|
24 in (61 cm) minimum clearance around left side and back
|
|
|
|
36 in (91.4 cm) minimum clearance at front of analyzer
|
|
|
Computer Workstation
|
24 in (61 cm) clearance in front
|
Sample Tube Barcode Specifications
|
|
Code 39
|
|
|
Code 128 (ISBT 128)
|
|
|
Interleaved 2 of 5
|
|
|
Codabar
|
Environmental Requirements
|
|
Ambient Temperature
|
15-25º C
|
|
|
Humidity
|
20-85%
|
|
|
 |

Electrical Requirements
|
|
Electrical Input
|
230 VAC, 50-60 Hz, single phase
|
|
|
Input Current
|
30 A circuit, dedicated circuit required
|
|
|
Maximum Power Fluctuation
|
208V AC +/- 10V
|
|
|
Electrical Connector
|
L6-30R (USA)
|
Environmental Output
|
|
Heat Dissipation
|
773 W (2637 BTU/Hr) during steady operation
|
|
|
Drain Output
|
None: Waste fluids drain to removable containers
|
NCCLS Water Requirements
Sample Throughput
|
|
3.5 hours to first results
|
|
|
100 samples processed per hour following the first result
|
Sample Capacity
|
|
180 tubes per sample carousel, in 9 sample racks
|
|
|
20 tubes per sample rack
|
Reagent Capacity
|
|
1 Reagent Kit per quadrant
|
|
|
4 quadrants on the Assay Reagent Carousel
|
|
|
Designed to accomodate menu expansion
|
|
|
1000 test on-board capacity
|
|
 |